Citation

Abstract

This article compares the presently used JPL hydrogen maser suspended atomic storage bulb to a new, rigid, single-plane mounted bulb. The new bulb incorporates three major design changes: (1) Mounting design. (2) Alterations to the collimator. (3) Decrease in mass. These design changes are expected to increase the long-term stability of the frequency standard by reducing its sensitivity to vibration and thermal effects.

Details

Volume
42-52
Published
August 15, 1979
Pages
21–25
File Size
748.3 KB