Citation

Abstract

A simple method for determination and correction of non-linearity in measurement systems is presented. The technique is applicable to a wide range of measuring systems. The basic concept, an analysis, and a sample application are given. Non-linearity of the Goldstone DSS 13 low noise HEMT 2.3-GHz (S-band) radiometer system results in noise temperature measurement errors. These errors are successfully corrected with this method.

Details

Volume
42-91
Published
November 15, 1987
Pages
57–66
File Size
325.9 KB